7 Mistakes to Avoid When Using pH Meters for testing beer brew batches

Thursday, October 26, 2023

pH meters have revolutionized the way brewers, chemists, scientists, and even kombucha makers measure acidity or alkalinity levels in their concoctions. While these devices are invaluable, they are also susceptible to errors, especially if not used or maintained correctly.

ph tester user mistakes to avoid

1. Electrode Storage Mishaps

Electrodes, the heart and soul of pH meters, are intricate devices that need meticulous care. One common error is storing the electrode dry. The science behind it is fascinating: pH electrodes contain three distinct glass layers. The outer and inner layers, both hydrated, are responsible for the electrode's sensitivity to pH changes. Dry storage dehydrates these layers, leading to inaccurate readings. The electrode should always be stored in a fresh storage solution to maintain its sensitivity. If by chance you've stored it dry, you can attempt to revive it by soaking it in a pH storage solution for several hours.

2. Cleaning Conundrums

Cleanliness is key for accurate pH readings. While it might seem logical to wipe the sensing glass to keep it clean, doing so can introduce static electricity, which interferes with readings. Instead, soak the probes in distilled or deionized water. On the flip side, neglecting to clean your electrode can lead to the formation of a 'hydrated layer' on the glass bulb, which also skews readings. Regular cleaning with the appropriate solution is essential.

3. Calibration Complications

Calibration is the cornerstone of accurate pH measurements. Infrequent or incorrect calibration can lead to grossly inaccurate readings. For those who use their pH meters daily, daily calibration might be the way to go. For less frequent users, calibrating before each use can ensure accurate readings.

4. Electrolyte Levels

Refillable electrodes have a distinct advantage: they allow users to replenish the electrolyte in the reference compartment. But if neglected, low electrolyte levels can lead to erratic readings. Regularly check and maintain the fill solution level to avoid this pitfall.

5. Water Storage Woes

Storing the electrode in deionized water is a cardinal sin in the world of pH meters. This type of water lacks ions, leading the electrode to lose its own ions in an attempt to achieve equilibrium. This ion drain renders the electrode useless over time. If this error has been made, a potential fix involves refilling the electrode with a potassium nitrate fill solution and then placing it in a storage solution.

6. The Age of the Electrode

Everything has a shelf life, including pH electrodes. Using an old, dried-up electrode is a recipe for inaccurate readings. It's best to replace old probes with new ones suited for the task at hand.

7. Partial Submersion

For those in a rush, partially submerging the probe might seem like a time-saving hack. In reality, it's a mistake. The entire pH sensing component and reference junction must be submerged in the solution to guarantee accurate readings.

In Conclusion

While pH meters are powerful tools, they require careful handling and maintenance to provide accurate results. For those who find this all too cumbersome, pH strips can offer a simpler albeit less precise alternative. For the dedicated, investing in a quality meter, like the Milwaukee MW102, can make the journey smoother. Remember, with great power (or in this case, precision) comes great responsibility!